🏠
📞  +353 1 6643768
 
📧  info@medaval.ie
Beurer BM 26
Device Model:
BM 26
Manufacturer:

OBL: Beurer GmbH, Söflinger Straße 218, 89077 Ulm, GERMANY.

OEM: Zhejiang Joytech Healthcare Co. Ltd., 365 Wuzhou Road, Hangzhou, Zhejiang 311100, CHINA.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
Germany, United Kingdom, Spain, France, Italy, Hungary and United States
Device Manual:
Device Specifications:
Description:
The Beurer BM 26 is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Beurer BM 26, to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP DHL (DE) General use (2015) Manufacturer declaration of equivalence
BP Stride BP Self-measurement for adults (Preferred) Equivalence claim without published evidence
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

H&L HL868BA

Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit. 2008 Oct;13(5):305-8. doi: 10.1097/MBP.0b013e32830d4bd0. PMID: 18799958.

ESH-IP:2002 - Pass General population

Sejoy Joytech DBP-1307 (Sejoy BP-1307, Joytech DBP-1307)

Lei L, Chen Y, Chen Q, Li Y, Wang JG. Validation of the SEJOY BP-1307 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2017 Dec;22(6):371-4. Epub: 2017 Aug 03. doi: 10.1097/MBP.0000000000000283. PMID: 28786853 .

ESH-IP:2010 - Pass General population