Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | Stride BP | Self-measurement for adults (Preferred) | Equivalence to the Yuwell YE660D (V2) claim without published evidence |
Legend: P Stride Equivalence
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Yuwell YE660D (V2)
Zhang HJ, Peng Y, Zhang J, Zhang J, Teng LN, Zhang SJ, Zhou DJ, Long MZ. Validation of the YuWell YE660D oscillometric upper-arm blood pressure monitor for clinic and home in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) and the International Organization for Standardization Amendment 1. 2020. Blood Press Monit. 2023 Oct 1;28(5):276-279. Epub: 2023 Jun 22. doi: 10.1097/MBP.0000000000000658. PMID: 37382148.
81060-2:2020 - Pass General population (Note: Two cuffs 22-32 cm (standard) and 22-45 cm (wide range))
Yuwell YE680A
Zhang P, Li X, Fang Z, Lu Y, Cui J, Du X, Hu R. Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: the Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B. Blood Press Monit. 2021 Dec 1;26(6):435-440. Epub: 2021 May 13. doi: 10.1097/MBP.0000000000000547. PMID: 34001755.
ESH-IP:2010 - Fail General population