Withings BPM Connect Pro (WPM06)
Device Name:
BPM Connect Pro
Device Model:
WPM06
Manufacturer:
Withings Europe (France), 2 Rue Maurice Hartmann, 92130 Issy-les-Moulineaux, FRANCE.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Connected Health Technologies:
Bluetooth
Availability:
Available Currently
Availability according to Countries or Regions:
United States
Description:
The Withings BPM Connect Pro (WPM06) is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Withings BPM Connect Pro (WPM06), to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Self-measurement and professional use for adults and pregnancy Manufacturer declaration of equivalence
BP Stride BP Self-measurement for adults and pregnancy (Preferred) Equivalence claim without published evidence
BP AMA VDL (US) Self-measurement for general use and pregnancy Equivalence claim without published evidence
Device Family:
Withings BPM Connect (WPM05)B,P, Withings BPM Connect Pro (WPM06)B,P
Legend: B BIHS Derivative, P Stride Equivalence
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Withings BPM Connect (WPM05)

Zelveian P, Topouchian J, Hakobyan Z, Asmar J, Gharibyan H, Asmar R. Clinical Accuracy of the Withings BPM Connect for Self-Blood Pressure Measurements in Pregnancy and Pre-Eclampsia: Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Vasc Health Risk Manag. 2022 Mar 26;18:181-189. doi: 10.2147/VHRM.S351313. PMID: 35370407. Available from: PMC8967988.

81060-2:2019 - Pass Normotensive pregnancy (n=15), Gestational hypertensive pregnancy (n=15), Pre-eclampsia (n=15)

Topouchian J, Zelveian P, Hakobyan Z, Gharibyan H, Asmar R. Accuracy of the Withings BPM Connect Device for Self-Blood Pressure Measurements in General Population - Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Vasc Health Risk Manag. 2022 Mar 31;18:191-200. doi: 10.2147/VHRM.S350006. PMID: 35386428. Available from: PMC8979570.

81060-2:2019 - Pass General population