📞  +353 1 6643768
 
📧  info@medaval.ie
TaiDoc TD-4209 Clever Chek
Device Name:
Clever Chek
Device Model:
TD-4209
Manufacturer:
TaiDoc Technology Corporation, 6F - 127 Wugong 2nd Road, 24888 Wugu District, New Taipei City, TAIWAN ROC.
Measuring functions:
Blood glucose
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Finger
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Description:
The TaiDoc TD-4209 Clever Chek is an automatic blood glucose meter. The accuracy of its blood glucose measurement technology has yet to be proven to MDR requirements. Blood glucose measurements are taken from the finger. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The technology used in the TaiDoc TD-4209 Clever Chek, to measure blood glucose, has failed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication. However, it was also assessed in another study, but the results have not been published and cannot be verified.
Recommendations:
Accuracy AssessmentRecommendationBasis
BG Medaval None While a clinical validation is claimed, it cannot be checked properly.
BG Institute of Health Economics (CA) General use No evidence provided
Validation Publications:

Freckmann G, Baumstark A, Jendrike N, Zschornack E, Kocher S, Tshiananga J, Heister F, Haug C. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol Ther. 2010 Mar;12(3):221-31. doi: 10.1089/dia.2009.0128. PMID: 20151773.

15197:2003 - Fail General population