📞  +353 1 6643768
 
📧  info@medaval.ie
Rossmax Accumed AF701
Device Name:
Accumed
Device Model:
AF701
Manufacturer:

Brand: Rossmax Swiss GmbH (Accumed), Tramstrasse 16, CH-9442 Berneck, SWITZERLAND.

Other: Rossmax International Ltd., 12F - 189 Kang Chien Road, Taipei 114, TAIWAN ROC.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Description:
The Rossmax Accumed AF701 is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
While the technology used in the Rossmax Accumed AF701, to measure blood pressure, has been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Self-measurement Manufacturer declaration of equivalence
BP ESH (Europe) Self-measurement The BIHS recommendation, despite the absence of the scientific evidence required by the same authors' criteria.
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Rossmax ME701

de Greeff A, Shennan AH. The Rossmax (ME 701 series) upper arm device: accuracy assessment in an adult population according to the International and the British Hypertension Society protocols. Blood Press Monit. 2008 Feb;13(1):43-8. doi: 10.1097/MBP.0b013e3282ca913f. PMID: 18199923.

BHS:1993 - Pass (A/A) General population

SP10:1996 - Pass General population