📞  +353 1 6643768
 
📧  info@medaval.ie
ReliOn Prime
Device Name:
Prime
Manufacturer:

OBL: Walmart Apollo LLC (ReliOn), 702 SW 8th Street, Bentonville, Arkasas 72716, UNITED STATES.

OEM: Arkray Global Business Inc., 1408 Koji, Konan cho, koka shi, Shiga, JAPAN.

Measuring functions:
Blood glucose
Primary Client Use:
Intended for both professional use and self-measurement
Measurement Site:
Finger and general Alternative Site Testing
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
United States
Device Manual:
Description:
The ReliOn Prime is an automatic blood glucose meter. Medaval has not found evidence proving the accuracy of its blood glucose measurement technology. Blood glucose measurements are normally taken from the finger but alternative site testing is allowed. It is intended for both professional use and self-measurement.
Assessment:
The technology used in the ReliOn Prime, to measure blood glucose, has failed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BG Medaval Not recommended This device has failed a clinical validation.
Consumer BodyYear(s)LevelComment
BG Consumer Reports (US) Latest: 2019-2022 Very Good
Validation Publications:

Klonoff DC, Parkes JL, Kovatchev BP, Kerr D, Bevier WC, Brazg RL, Christiansen M, Bailey TS, Nichols JH, Kohn MA. Investigation of the Accuracy of 18 Marketed Blood Glucose Monitors. Diabetes Care. 2018 Aug;41(8):1681-8. Epub: 2018 Jun 13. doi: 10.2337/dc17-1960. PMID: 29898901. Available from: www.diabetestechnology.org.

DTS:2016 - Fail General population