📞  +353 1 6643768
 
📧  info@medaval.ie
Philips IntelliVue X2 (M3002A)
Device Name:
IntelliVue X2
Device Model:
M3002A
Manufacturer:
Philips Electronics North America Corp., 3000 Minuteman Road, M/S 109, Andover, MA 01810, UNITED STATES.
Measuring functions:
Blood pressure, SpO2, ECG, Respiration
Primary Client Use:
Intended for patient monitoring
Measurement Site:
BP: Arm and Intra-Arterial; SpO2: Finger
Measurement Occurrence:
BP: Single, intermittent and continuous measurements; SpO2: Continuous measurement
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Device Manual:
Description:
The Philips IntelliVue X2 (M3002A) is a patient monitor. Although its blood pressure measurement technology has been evaluated, it is not possible to base a determination on its accuracy from this research and information regarding the accuracy of its oxygen saturation measurement technology does not appear to be available. Both upper arm and intra-arterial measurements can be taken and oxygen saturation measurements are taken from the finger. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Philips IntelliVue X2 (M3002A), to measure blood pressure, has been assessed, in a specific population, according to a non-standard protocol. There appears to be no peer-reviewed clinical validation information available on the technology used in the Philips IntelliVue X2 (M3002A) to measure peripheral oxygen saturation.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol.
BP Japanese Society of Hypertension Professional use (2016-2018) No evidence provided
Validation Publications:

Feenstra RK, Allaart CP, Berkelmans GFN, Westerhof BE, Smulders YM. Accuracy of oscillometric blood pressure measurement in atrial fibrillation. Blood Press Monit. 2018 Apr;23(2):59-63. Epub: 2017 Dec 17. doi: 10.1097/MBP.0000000000000305. PMID: 29256921.

Ad Hoc protocol Atrial fibrillation (n=23)