📞  +353 1 6643768
 
📧  info@medaval.ie
Hillrom Home 1700 (H-BP100SBP)
Device Name:
Home 1700
Device Model:
H-BP100SBP
Manufacturer:

OBL: Hill-Rom Holdings Inc., Suite 1000, 130 East Randolph Street, Chicago, IL 60601, UNITED STATES.

Former: Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153, UNITED STATES.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
United States
Device Manual:
Description:
The Hillrom Home 1700 (H-BP100SBP) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Hillrom Home 1700 (H-BP100SBP), to measure blood pressure, has passed in a clinical validation study, in a specific population, according to a recognised standard protocol, as published in a peer-reviewed publication. It has also been declared as being substantially equivalent to that used in another device, according to 510(k), but this is not compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommended in certain condition(s) Validation in specific condition(s). See publications for details.
BP Stride BP Self-measurement for adults (Preferred) Published evidence
BP AMA VDL (US) Self-measurement Published evidence plus 510(k) K161886
Consumer BodyYear(s)LevelComment
BP Amazon Best Sellers (US) Latest: 2019 A (3.8 / 5)
Device Family:
Validation Publications:

Alpert BS. Validation of the Welch Allyn Home blood pressure monitor with professional SureBP algorithm with a special feature of accuracy during involuntary (tremor) patient movement. Blood Press Monit. 2019 Apr;24(2):89-92. Epub: 2019 Feb 4. doi: 10.1097/MBP.0000000000000365. PMID: 30724792.

81060-2:2013 - Pass Adult subjects whose oscillometric pressures were augmented with simulator-produced tremors. (n=85)

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Hillrom ProBP 2000

Alpert BS. Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing. Blood Press Monit. 2018 Dec;23(6):315-317. Epub: 2018 Oct 11. doi: 10.1097/MBP.0000000000000350. PMID: 30312180.

81060-2:2013 - Pass Adults (n=53) and children (n=35)

Transtek TMB-1776

Liu Z, Chen L. Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults. Blood Press Monit. 2019 Dec;24(6):319-322. Epub: 2019 Sep 26. doi: 10.1097/MBP.0000000000000406. PMID: 31567185.

ESH-IP:2010 - Pass General population

Transtek TMB-986

Liu WJ, Li SG, Song Z, Gong W. Validation of the TRANSTEK blood pressure monitor TMB-986 for home blood pressure monitoring according to the International Protocol. Blood Press Monit. 2010 Oct;15(5):278-80. doi: 10.1097/MBP.0b013e32833e43ca. PMID: 20729726.

ESH-IP:2002 - Pass General population