Ergoline Ergoscan
Device Name:
Ergoscan
Manufacturer:

OBL: Ergoline GmbH, Lindenstraße 5, 72475 Bitz, GERMANY.

OEM: EnviteC-Wismar GmbH, Alter Holzhafen 18, 23966 Wismar, GERMANY.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for ambulatory blood pressure monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Available Currently
Availability according to Countries or Regions:
Ireland
Device Manual:
Description:
The Ergoline Ergoscan is an ambulatory blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for ambulatory blood pressure monitoring.
Assessment:
While the technology used in the Ergoline Ergoscan, to measure blood pressure, has been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) ABPM Manufacturer declaration of equivalence
BP ESH (Europe) ABPM The BIHS recommendation, despite the absence of the scientific evidence required by the same authors' criteria.
Device Family:
Ergoline ErgoscanB, EnviteC PhysioQuant ABPM (45-00-0500)B
Legend: B BIHS Derivative
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

EnviteC PhysioQuant ABPM (45-00-0500)

Langewitz W, Tanner S. Validation of the PhysioQuant blood pressure measuring monitor in adults. Blood Press Monit. 2009 Oct;14(5):220-1. doi: 10.1097/MBP.0b013e328330c830. PMID: 19704372.

ESH-IP:2002 - Pass General population