📞  +353 1 6643768
 
📧  info@medaval.ie
Dinamap ProCare 200
Device Name:
ProCare 200
Manufacturer:
GE Medical Systems Ltd. (Dinamap), Amersham Place, Little Chalfont, Buckinghamshire HP7 9NA, UNITED KINGDOM.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for patient monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Available Currently
Device Manual:
Description:
The Dinamap ProCare 200 is a patient monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Dinamap ProCare 200, to measure blood pressure, has failed in a clinical validation study, in a specific population, according to a recognised standard protocol, as published in a peer-reviewed publication. While this technology has also been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Professional use Manufacturer declaration of equivalence
Device Family:
GE Carescape V100B, Dinamap Procare 420B, Dinamap Carescape V100B, Dinamap ProCare 100B,P, Dinamap ProCare 200B, Dinamap ProCare (DPC 100X-EN)P
Legend: B BIHS Derivative, P Stride Equivalence
Validation Publications:

Lee CG, Park HM, Shin HJ, Moon JS, Hong YM, Kim NS, Ha IS, Chang MJ, Oh KW. Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents. Korean J Pediatr. 2011 Nov;54(11):463-9. Epub: 2011 Nov 30. doi: 10.3345/kjp.2011.54.11.463. PMID: 22253643. Available from: PMC3254892.

ESH-IP:2010 - Fail Children and adolescents (n=45)

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Dinamap ProCare (DPC 100X-EN)

de Greeff A, Reggiori F, Shennan AH. Clinical assessment of the DINAMAP ProCare monitor in an adult population according to the British Hypertension Society Protocol. Blood Press Monit. 2007 Feb;12(1):51-55. doi: 10.1097/MBP.0b013e3280858b73. PMID: 17303988.

BHS:1993 - Pass (A/A) General population

SP10:1992 - Pass General population

Dinamap ProCare 100

Reinders A, Reggiori F, Shennan AH. Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population. Blood Press Monit. 2006 Oct;11(5):293-6. PMID: 16932038.

ESH-IP:2002 - Pass General population