🏠
📞  +353 1 6643768
 
📧  info@medaval.ie
A&D UA-851PBT-C
Device Model:
UA-851PBT-C
Manufacturer:
A&D Company Ltd., 3-23-14 Higashi-Ikebukuro, Toshima-Ku, 170-0013 Tokyo, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Description:
The A&D UA-851PBT-C is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for professional use.
Assessment:
The technology used in the A&D UA-851PBT-C, to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP Japanese Society of Hypertension Self-measurement (2016-2022) Implied equivalence to the A&D UA-851, UA-852, UA-853, UA-854 or UA-855
Device Family:
A&D UA-851B,J, A&D UA-852B,J, A&D UA-853B,J, A&D UA-854B,J, A&D UA-855B,J, A&D UA-851PBT-CJ
Legend: B BIHS Derivative, J JHS Implied Equivalence
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

A&D UA-851

Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.

ESH-IP:2002 - Pass General population

SP10:1987 - Pass General population

A&D UA-852

Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.

ESH-IP:2002 - Pass General population

SP10:1987 - Pass General population

A&D UA-853

Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.

ESH-IP:2002 - Pass General population

SP10:1987 - Pass General population

A&D UA-854

Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.

ESH-IP:2002 - Pass General population

SP10:1987 - Pass General population

A&D UA-855

Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.

ESH-IP:2002 - Pass General population

SP10:1987 - Pass General population