Device Model:
UA-1120ECO
Manufacturer:
A&D Company Ltd., 3-23-14 Higashi-Ikebukuro, Toshima-Ku, 170-0013 Tokyo, JAPAN.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The A&D UA-1120ECO is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the A&D UA-1120ECO, to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP Japanese Society of Hypertension Self-measurement (2025) Manufacturer declaration of equivalence to the A&D UA-1100NFC
Device Family:
A&D UA-1200CNBLE - 1200 UltraConnectZ, A&D UA-1200BLEJ, A&D UA-1100NFCJ, A&D UA-1120ECOJ, A&D UA-1300BLEJ, Hingmed DBP-01HPF, Hingmed DBP-01PF
Legend: J JHS Implied Equivalence, F FDA 510(k), Z Almost Certain
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

A&D UA-1100NFC

Kario K, Tomitani N, Hoshide S. Validation of the A&D BP UA-1100NFC/UA-1100NFC-W, hoseless upper arm-type home blood pressure devices, according to the ISO81060-2:2018/Amd 1:2020 protocol. Blood Press Monit. 2025 Jun 1;30(3):130-135. Epub: 2025 Jan 8. doi: 10.1097/MBP.0000000000000741. PMID: 39831763.

81060-2:2018/Amd 1:2020 - Pass General population

A&D UA-1200BLE

Kario K, Saito K, Sato K, Hamasaki H, Suwa H, Okura A, Hoshide S. Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2: 2013 standard. Blood Press Monit. 2018 Dec;23(6):312-314. Epub: 2018 Sep 7. doi: 10.1097/MBP.0000000000000347. PMID: 30199378.

81060-2:2013 - Pass General population

Hingmed DBP-01P

Fania C, Giletto A, Zanello V, Palatini P. Validation of the Hingmed DBP-01P clinical automatic blood pressure monitor provided with wide-range cuff, evaluated in a general population according to the ISO 81060-2:2018 protocol. Blood Press Monit. 2023 Apr 1;28(2):116-120. Epub: 2023 Feb 17. doi: 10.1097/MBP.0000000000000638. PMID: 36916471.

81060-2:2018 - Pass General population