Yuwell YE666CR (V2)
Device Model:
YE666CR (Note: 2 of 2)
Manufacturer:
Jiangsu Yuyue Medical Equipment and Supply Co. Ltd. (Yuwell), Yunyang Industrial Park, Danyang, Jiangsu 212300, CHINA.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The Yuwell YE666CR (V2) is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Yuwell YE666CR (V2), to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP Stride BP Self-measurement for adults (Preferred) Equivalence to the Yuwell YE660D (V2) claim without published evidence
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Yuwell YE660D (V2)

Zhang HJ, Peng Y, Zhang J, Zhang J, Teng LN, Zhang SJ, Zhou DJ, Long MZ. Validation of the YuWell YE660D oscillometric upper-arm blood pressure monitor for clinic and home in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) and the International Organization for Standardization Amendment 1. 2020. Blood Press Monit. 2023 Oct 1;28(5):276-279. Epub: 2023 Jun 22. doi: 10.1097/MBP.0000000000000658. PMID: 37382148.

81060-2:2020 - Pass General population (Note: Two cuffs 22-32 cm (standard) and 22-45 cm (wide range))