OBL: MED TRUST Handelsgesellschaft m.b.H. (Wellion), Gewerbepark 10, A-7221 Marz, AUSTRIA.
OEM: AgaMatrix Inc., 7C Raymond Avenue, Salem, NH 03079, UNITED STATES.
Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | Legacy approval | Older clinical validation; older protocol |
Freckmann G, Baumstark A, Jendrike N, Zschornack E, Kocher S, Tshiananga J, Heister F, Haug C. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol Ther. 2010 Mar;12(3):221-31. doi: 10.1089/dia.2009.0128. PMID: 20151773.
15197:2003 - Pass General population
Chlup R, Doubravova B, Peterson K, Zapletalova J, Bartek J. Wavesense technology glucometer Linus for routine self-monitoring and clinical practice. Acta Diabetol. 2011 Mar;48(1):35-40. Epub: 2010 Aug 9. doi: 10.1007/s00592-010-0213-5. PMID: 20694824.
Ad Hoc protocol General population
Abstract concludes with "Glucometer Linus readings correspond to the ISO and FDA standards. So, Linus appears to be an accurate device for PG-self-monitoring and clinical practice." However, it is unclear which standards are used.
Chlup R, Doubravova B, Bartek J, Zapletalova J, Krystynik O, Prochazka V. Effective assessment of diabetes control using personal glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA). Dis Markers. 2013;35(6):895-905. PMID: 24427777.
Ad Hoc protocol General population
While ISO 15197:2003 and ISO 15197:2013 are mentioned, it is unclear if either was folowed.