Welch Allyn 6000 Series NIBP Module 103386
Device Name:
6000 Series NIBP Module
Device Model:
103386
Manufacturer:
OBL: Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153, UNITED STATES.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Module intended as a component of a patient monitor
Measurement Site:
Upper Arm
Measurement Occurrence:
Single and intermittent measurements
Availability:
Discontinued but still available
Description:
The Welch Allyn 6000 Series NIBP Module 103386 is a blood pressure module for a patient monitor. Medaval has not found evidence proving the accuracy of its blood pressure measurement technology. Blood pressure measurements are taken from the upper arm. It is intended for use in a bedside patient monitor. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
There appears to be no peer-reviewed clinical validation information available on the technology used in the Welch Allyn 6000 Series NIBP Module 103386 to measure blood pressure.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Not recommended This device has not been clinically validated.