Novacor DiaSys 200R
Device Name:
DiaSys 200R
Manufacturer:
Novacor SAS, 4 passage Saint-Antoine, 92508 Rueil-Malmaison, Cedex, FRANCE.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for ambulatory blood pressure monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Obsolete
Description:
The Novacor DiaSys 200R is an ambulatory blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate. Blood pressure measurements are taken from the upper arm. It is intended for ambulatory blood pressure monitoring. This device is now obsolete.
Assessment:
The technology used in the Novacor DiaSys 200R, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Legacy approval Older clinical validation; older protocol
Validation Publications:

O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens. 1991;9(Suppl 5):S9-15.

BHS:1990 - Fail (C/C) General population

SP10:1987 - Pass General population

Barthélémy JC, Geyssant A, Auboyer C, Antoniadis A, Berruyer J, Lacour JR. Accuracy of ambulatory blood pressure determination: a comparative study. Scand J Clin Lab Invest. 1991 Sep;51(5):461-6. PMID: 1947731.

IA protocol General population

Mora-Maciá J, Ocón Pujadas JM, Doñate Cubells T, del Río Pérez G. Validación de un monitor no invasivo de registro de presión arterial ambulatoria: El DIASYS 200 Novacor. [The evaluation of a noninvasive monitor to record ambulatory arterial pressure: the DIASYS 200 Novacor] Med Clin (Barc). 1993 Oct 16;101(12):450-4. PMID: 8231367. In Spanish.

SP10:1987 - Fail General population (Note: Fail / Pass)

Relevant Publications:

O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207, Novacor DIASYS 200, Del Mar Avionics Pressurometer IV and Takeda TM-2420 ambulatory systems according to British and American criteria. J Hypertens. 1991 Dec;9(Suppl 6):S332-3. PMID: 1818987.

This is a supplementary presentation of validation results for the Novacor Diasys 200R.

See: O'Brien E et al. 1991

O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens. 1991 Jun;9(6):569-70. PMID: 1653297. (Details from: journals.lww.com).

This is a supplementary presentation of validation results for the Novacor Diasys 200R.

See: O'Brien E et al. 1991

O'Brien E, Atkins N, Mee F, O'Malley K. Comparative accuracy of six ambulatory devices according to blood pressure levels. J Hypertens. 1993 Jun;11(6):673-5. PMID: 8397247.

This is a supplementary presentation of validation results for the Novacor Diasys 200R.

See: O'Brien E et al. 1991