| Accuracy Assessment | Recommendation | Basis | |
| BP | Medaval | ★★ Recommendation | Recent clinical validation; recent protocol |
| BP | Stride BP | Self-measurement for adults and diabetes mellitus (Preferred) | Internal report plus equivalence claim |
Legend: B BIHS Derivative, P Stride Equivalence, Z Almost Certain
Menti A, Ntousopoulos V, Lempidaki P, Bakalakou KR, Komnianou A, Mariglis D, Ntaskagianni E, Kollias A, Stergiou GS. Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KV1-5X (BP B6 Connect) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018). Blood Press Monit. 2026 Feb 1;31(1):35-38. Epub: 2025 Sep 10. doi: 10.1097/MBP.0000000000000771. PMID: 40931782.
81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population
Hansen KW. Comments on 'Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KV1-5X (BP B6 Connect)...'. Blood Press Monit. 2026 Apr 1;31(2):101. Epub: 2026 Feb 25. doi: 10.1097/MBP.0000000000000784. PMID: 41746602.
The publication is a letter or correspondence in which the Microlife BP B6 Connect (BP 3KV1-5X) is referenced.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Microlife BP B3 AFib (BP 3KT1-3N)
Beime B, Bramlage C, Krüger R, Deutsch C, van Mark G, Bramlage P, Botta B. Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2: 2019 protocol. Blood Press Monit. 2021 Aug 1;26(4):299-304. Epub: 2021 Mar 15. doi: 10.1097/MBP.0000000000000530. PMID: 33734119.
81060-2:2018 - Pass General population