Brand: Huawei Technologies Co. Ltd., Bantian Huawei Base, Longgang District, Shenzhen, Guangdong 518129, CHINA.
Other: Huawei Device Co. Ltd., 2 Xincheng Road, Songshan Lake Zone, Dongguan, Guangdong 523808, CHINA.
| Accuracy Assessment | Recommendation | Basis | |
| BP | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
| BP | Stride BP | Self-measurement for adults | Published evidence |
Zhou Y, Zhang W, Wang XY, Zhou Y, Li Y, Wang JG. Accuracy of a wearable watch-type oscillometric blood pressure monitor in rest and ambulatory blood pressure measurements. Hypertens Res. 2025 Nov;48(11):2961-2968. Epub: 2025 Aug 27. doi: 10.1038/s41440-025-02345-2. PMID: 40866549.
81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population (Note: Primary 85-subject study)
81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population (Note: Stress study, as required for ABPM devices (n=35))
81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population (Note: Ambulatory study (n=35))
Liu Y, Lu Y, Wu J, Zhang T, Wang B, Tian Z. Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation Study. J Clin Hypertens (Greenwich). 2025 Nov;27(11):e70183. doi: 10.1111/jch.70183. PMID: 41263383.
81060-2:2018 - Pass General population (Note: Seated position)
81060-2:2018 - Pass General population (Note: Supine position)
Ad Hoc protocol General population (Note: During sleep)
Seoane-Casqueiro Ó, Vázquez-Agra N, Cruces-Sande A, López-Prado E, López-Paz JE, Camafort M, Casariego-Vales E, Hermida-Ameijeiras Á, Pose-Reino A. Clinical reliability of a wrist-worn oscillometric ambulatory blood pressure monitor: an independent agreement evaluation. Rev Clin Esp (Barc). Epub: 2026 Mar 23:502523 p. doi: 10.1016/j.rceng.2026.502523. PMID: 41881377.
Measurements from the Huawei Watch D2 (LCA-B11) are compared to those from the SunTech Oscar 2, though not in a formal validation.