| Accuracy Assessment | Recommendation | Basis | |
| BG | Medaval | None | When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol. |
Wadwa RP, Laffel LM, Shah VN, Garg SK. Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with Diabetes. Diabetes Technol Ther. 2018 Jun;20(6):395-402. Epub: 2018 Jun 14. doi: 10.1089/dia20180150. PMID: 29901421. Available from: PMC6110124.
Ad Hoc protocol Adults and children (Note: The new factory-calibrated G6 real-time CGM system provides accurate readings for 10 days and removes several clinical barriers to broader CGM adoption.)
Huang E, Demirel S, Bliss C, Savaser D, Castle JR. Reliability of the Dexcom G6 Continuous Glucose Monitor During Hyperbaric Oxygen Exposure. Diabetes Technol Ther. 2020 May;22(5):360-366. Epub: 2020 Feb 6. doi: 10.1089/dia20190390. PMID: 31916854. Available from: PMC7196364.
In those without DM, there is a statistically significant, but clinically insignificant, effect of gas+pressure condition on the glucose estimate detected by the Dexcom G6 CGM system. (n=26)
Huang E, Demirel S, Bliss C, Savaser D, Castle JR. Reliability of the Dexcom G6 Continuous Glucose Monitor During Hyperbaric Oxygen Exposure. Diabetes Technol Ther. 2020 May;22(5):360-366. Epub: 2020 Feb 6. doi: 10.1089/dia20190390. PMID: 31916854. Available from: PMC7196364.
In those without DM, there is a statistically significant, but clinically insignificant, effect of gas+pressure condition on the glucose estimate detected by the Dexcom G6 CGM system. (n=26)
Herzig
Cardiac surgery using hypothermic extracorporeal circulation challenges the accuracy of the Dexcom G6 CGM although recovery appears to occur thereafter.
Avari P, Tang W, Jugnee N, Hersi I, Al-Balah A, Tan T, Frankel AH, Oliver N, Reddy M. The Accuracy of Continuous Glucose Sensors in People with Diabetes Undergoing Hemodialysis (ALPHA Study). Diabetes Technol Ther. 2023 Jul;25(7):447-456. doi: 10.1089/dia.2023.0013. PMID: 36961385.
Further studies are required to evaluate Dexcom G6 accuracy in people on hemodialysis.