Del Mar Avionics Pressurometer IV
Device Name:
Pressurometer IV
Manufacturer:
Del Mar Avionics Inc., 11-B Marconi, Irvine, CA 92618, UNITED STATES.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for ambulatory blood pressure monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Obsolete
Description:
The Del Mar Avionics Pressurometer IV is an ambulatory blood pressure monitor. Although its blood pressure measurement technology has been evaluated, it is not possible to base a determination on its accuracy from this research. Blood pressure measurements are taken from the upper arm. It is intended for ambulatory blood pressure monitoring. This device is now obsolete.
Assessment:
The technology used in the Del Mar Avionics Pressurometer IV, to measure blood pressure, has failed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication. It has also been assessed, in a general population, according to a non-standard protocol.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol.
Validation Publications:

Santucci S, Cates EM, James GD, Schussel YR, Steiner D, Pickering TG. A comparison of two ambulatory blood pressure monitors, the Del Mar Avionics Pressurometer IV and the Spacelabs 90202. Am J Hypertens. 1989 Oct;2(10):797-9. PMID: 2803673.

Ad Hoc protocol General population (Note: (n=48))

Barthélémy JC, Geyssant A, Auboyer C, Antoniadis A, Berruyer J, Lacour JR. Accuracy of ambulatory blood pressure determination: a comparative study. Scand J Clin Lab Invest. 1991 Sep;51(5):461-6. PMID: 1947731.

IA protocol General population

O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens. 1991;9(Suppl 5):S1-S7.

BHS:1990 - Fail (C/D) General population

Relevant Publications:

White WB, Lund-Johansen P, Omvik P. Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exercise. Am J Cardiol. 1990 Jan 1;65(1):60-66. doi: 10.1016/0002-9149(90)90026-w. PMID: 2294682.

At rest, non-invasive ambulatory BP monitors are similar in accuracy to clinicians in BP determination but during exercise there are differences among the monitors that make some of them unsuitable for measurements while in motion.

O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the SpaceLabs 90207, Novacor DIASYS 200, Del Mar Avionics Pressurometer IV and Takeda TM-2420 ambulatory systems according to British and American criteria. J Hypertens. 1991 Dec;9(Suppl 6):S332-3. PMID: 1818987.

This is a supplementary presentation of validation results for the Del Mar Avionics Pressurometer IV.

See: O'Brien E et al. 1991

O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens. 1991 Jun;9(6):567-8. PMID: 1653296. (Details from: journals.lww.com).

This is a supplementary presentation of validation results for the Del Mar Avionics Pressurometer IV.

See: O'Brien E et al. 1991

O'Brien E, Atkins N, Mee F, O'Malley K. Comparative accuracy of six ambulatory devices according to blood pressure levels. J Hypertens. 1993 Jun;11(6):673-5. PMID: 8397247.

This is a supplementary presentation of validation results for the Del Mar Avionics Pressurometer IV.

See: O'Brien E et al. 1991