Citizen CHUA715
Device Model:
CHUA715
Manufacturer:
Citizen Systems Japan Co. Ltd., 6-1-12 Tanashi-cho, Nishi-Tokyo-Shi, Tokyo 188-8511, JAPAN.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The Citizen CHUA715 is an automatic blood pressure monitor. Medaval has not found evidence proving the accuracy of its blood pressure measurement technology. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
While the technology used in the Citizen CHUA715, to measure blood pressure, has been declared as being equivalent to that used in another device, not only has no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4, but there is no evidence provided on a validation of the other device.
Recommendations:
| Accuracy Assessment | Recommendation | Basis | |
| BP | Medaval | Not recommended | This device has not been shown to have been clinically validated. |
| BP | Japanese Society of Hypertension | Self-measurement (2023-2024) | Implied equivalence claim, without published evidence, to a device that does not apper to exist, the |
Device Family:
Citizen CHUA516J, Citizen CHUA715J, Citizen CHUA716-BKJ, Citizen CHUH 533J, Citizen CHUH 719J, Citizen CHUH904CJ
Legend: J JHS Implied Equivalence