ChoiceMMed MD300C23
Device Model:
MD300C23
Manufacturer:
Beijing Choice Electronic Technology Co. Ltd. (ChoiceMMed), 4F Jingyang Science and Technology Building, 15 Xijing Road, Shijingshan District, Beijing 100041, CHINA.
Measuring functions:
SpO2 and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Finger
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The ChoiceMMed MD300C23 is a fingertip automatic pulse oximeter. Although its oxygen saturation measurement technology has been evaluated, it is not possible to base a determination on its accuracy from this research. Oxygen saturation measurements are taken from the finger. It is intended for self-measurement and home use.
Assessment:
The technology used in the ChoiceMMed MD300C23, to measure peripheral oxygen saturation, has been assessed, in a general population, according to a non-standard protocol.
Recommendations:
Accuracy AssessmentRecommendationBasis
SpO2 Medaval None When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol.
Validation Publications:

Lipnick MS, Feiner JR, Au P, Bernstein M, Bickler PE. The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health Implications. Anesth Analg. 2016 Aug;123(2):338-45. doi: 10.1213/ANE.0000000000001300. PMID: 27089002.

Ad Hoc protocol General population (Note: Modified ISO 80601-2-61:2011 study)