Bionime Rightest GM210
Device Name:
Rightest
Device Model:
GM210
Manufacturer:
Bionime Corporation, No 100 - Sec. 2, Daqing Street, South District, Taichung City 40242, TAIWAN ROC.
Measuring functions:
Blood glucose
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Finger
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The Bionime Rightest GM210 is an automatic blood glucose meter. Medaval has not found evidence proving the accuracy of its blood glucose measurement technology. Blood glucose measurements are taken from the finger. It is intended for self-measurement and home use.
Assessment:
The technology used in the Bionime Rightest GM210, to measure blood glucose, has failed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BG Medaval Not recommended This device has failed a clinical validation.
Validation Publications:

Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.

15197:2003 - Fail General population

15197:2013 (aka 15197:2015) - Fail General population