AXD AXD-821
Device Model:
AXD-821
Manufacturer:
Guangzhou AXD Electronic Co. Ltd., Factory 3, 9 Changxin Road, Changshapu, Zhongluotan Town, 510550 Baiyun District, Guangzhou City, Guangdong Province, CHINA.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The AXD AXD-821 is an automatic blood pressure monitor. Medaval has not found evidence proving the accuracy of its blood pressure measurement technology. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
There appears to be no peer-reviewed clinical validation information available on the technology used in the AXD AXD-821 to measure blood pressure. However, this device appears to be very similar to another device by the same manufacturer or a device within a family of equivalent devices. At least one validation needs to be published and equivalence needs to be proven according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Not recommended This device has not been shown to have been clinically validated.
Device Family:
AXD AXD-815Z, AXD AXD-821Z
Legend: Z Almost Certain